新势力光电供应美国Wilks公司的InfraCal Soot Meter烟尘仪,是一种ASTM 7686-11认可的用于测量柴油发动机润滑油的碳(煤烟)含量的测量方法。该系列烟尘分析仪具有如下特点:使用方便,检测速度快、测量范围大、无需样品制备和稀释过程,广泛用于润滑剂的评估和筛选,应用场所包括:油品检测实验室、发动机测试实验室、车站维修点、建筑或采矿点、航船、发电站。

InfraCal Soot Meter
Model HATR-CP
Analytical Wavelength
3.9μm, 2553cm-1
Sample Surface
Cubic Zirconia ATR Crystal
Scratch resistant, impervious to acids
Display
4 digit, 7-segment red LED, 5/8 in. character height
Power Requirements
12V DC, 2.0A, 12V battery pack or adapter available
Power Supply
Universal AC/DC (supplied as standard)
Weight
2.0kgs
Dimensions
165×165×225mm3
Suggested Operating Range
4°C-45°C
Typical Measurement Range
0-15% Soot
Output Resolution
0.0001
Analysis Time
Less than 30 seconds
Communication Port
RS232 port, 9-pin D-sub Female, PC compatible
Instrument Repeatability
0-3% soot ±0.1%, 3-15% soot ±0.3%
Non-volatile Memory Calibration
Factory: 0-10% soot
User: Up to 20 point curve fitting calibration, plus zero

Part Number
Descriptions
405-0011
InfraCal Soot Meter Plus, Model HATR-CP. Complete with HATR Sample Stage for direct measurement of suspended carbon (soot) level in diesel engine oils (factory calibrated to 10% soot concentration), Power Supply and Instruction Manual.
Accessories
403-0025
Set of 6 Reference Oils in dropper bottles, appropriate for calibration and verification of reproducibility 0-12% soot range of Model HATR-CP and Model HATR-C. This set of reference oils has been correlated with the results obtained via thermal gravimetric analysis (TGA) of fresh used diesel oil samples.
403-0018
InfraLog Serial Printer Prints calibration and measurement data, complete with sequence number, date and time. Includes all required cables.
403-0013
Carrying Case with pre-diced pluck foam, 470×370×203mm3, 3.6kgs.
403-0003
External 12V Battery Pack with 120V/60Hz charger (for use in North America). Includes carrying case with shoulder strap and cigarette lighter adapter cable for connection to analyzer. Provides 16 hours of continuous operation.
403-0002
Cigarette Lighter Adapter Cable
403-0005
Dust Cover
403-0001
Universal Switching Power Supply and Line Cord for use with InfraCal Analyzers. 90-264VAC, 50-60Hz input, regulated 12V/25W output
403-0014
Replacement Source/Cable Assembly
100-1005
InfraCal Soot Meter User's Manual
100-1011
InfraCal Soot Meter Plus User's Manual

Comparative Soot Measurements On Used Diesel Oil Samples
FT-IR, TGA, and InfraCal Soot Meter
Provided by Mobil Oil Company, Paulsboro Technical Center
Field Test Samples
Lab Engine Test Samples
Soot, % mass
Soot, % mass
Sample ID
FT-IR
TGA
Soot Meter 1
Soot Meter 2
Engine/Hours
TGA
Soot Meter 1
Soot Meter 2
A1
1.46
 
1.4
1.4
Mack T-8
 
 
 
B1
1.59
 
1.6
1.6
Hours:
0
0.13
0.1
0.0
B2
1.82
 
1.8
1.9
 
50
1.76
1.7
1.7
B3
1.94
 
1.9
2.0
 
100
3.13
3.2
3.1
B4
2.56
 
2.5
2.5
 
150
4.48
4.5
4.4
A2
2.74
2.65
2.6
2.6
 
175
5.16
5.0
5.0
B5
2.68
 
2.7
2.7
 
200
5.80
5.8
5.8
B6
2.81
2.76
2.8
2.9
 
225
6.38
6.1
6.0
B7
2.82
2.90
2.6
2.7
 
250
6.92
6.1
6.2
B8
 
3.15
3.0
3.1
DDC Series 60
 
 
 
B9
 
3.21
3.2
3.2
Hours:
0
0.09
0.0
0.0
B10
 
3.53
3.3
3.1
 
24
0.83
0.8
0.8
A3
 
3.54
3.5
3.6
 
48
1.40
1.4
1.4
A4
 
4.05
4.0
4.0
 
72
1.98
1.9
1.9
A5
 
4.40
4.3
4.4
 
96
2.52
2.5
2.5
A6
 
4.99
4.5
4.6
 
120
3.05
3.0
2.9
A7
 
6.12
5.1
5.1
 
144
3.58
3.5
3.5
A8
 
6.47
5.7
5.7
 
168
4.10
4.0
4.1
A9
 
7.00
6.0
6.1
 
192
4.57
4.5
4.6
 
 
 
 
 
 
216
5.03
4.8
4.9
 
 
 
 
 
 
240
5.33
5.0
5.0
Note: A & B series samples are different oil formulations.