来宝网Logo

热门词:生物显微镜 水质分析仪 微波消解 荧光定量PCR 电化学工作站 生物安全柜

现在位置首页>自动化仪器仪表>流量仪表>流量计> FLT-39C_FLT-39T_FLT-40G_FLT-40SA_FLT-40ST_FLT-40C_

商品编号:74797
分享

FLT-39C_FLT-39T_FLT-40G_FLT-40SA_FLT-40ST_FLT-40C_

价    格:询价

产    地:美国更新时间:2024/11/12 16:14:13

品    牌:欧米茄型    号:FLT-39C

状    态:正常点击量:342

15221105376
联系我时,请说明是在来宝网上看到的,谢谢!

上海茂培电子科技有限公司

联 系 人:范先生

电     话:15221105376

传     真:

等     级: (第 10年)

性     质:贸易型,

联系我时请说在来宝网上看到的,谢谢!




FLT-39C_FLT-39T_FLT-40G_FLT-40SA_FLT-40ST_FLT-40C_FLT-40T流量计OMEGA

多管组件


气体比例调节器


本页所示的流量管可安装在FL3000、气体比例FL-1GP系列或多管FL-2AA系列150 mm转子流量计中。所示流量基于21°C(70°F)和14.7 psia下空气和水的校准。对于其他流体或条件,以下方程提供粗略的近似值。



产品参数

FLT-39C_FLT-39T_FLT-40G_FLT-40SA_FLT-40ST_FLT-40C_FLT-40T流量计OMEGA

where:

Qw = equivalent water flow capacity at Standard Conditions

QL = max flow of metered liquid

df = density of float selected

   Glass: 2.53 g/ml

   Sapphire: 3.98 g/ml

   Stainless Steel: 8.04 g/ml

   Carboloy: 14.98 g/ml

   Tantalum: 16.58 g/ml

dL = density of metered liquid, g/ml

dw = density of water at Standard Conditions (1.0 g/ml)

(note: does not include viscosity effect)



产品介绍

FLT-39C_FLT-39T_FLT-40G_FLT-40SA_FLT-40ST_FLT-40C_FLT-40T流量计OMEGA

Multitube Assemblies

Gas Proportioners

The flow tubes shown on this page can be installed into the FL3000, gas proportioning FL-1GP Series or multiple tube FL-2AA Series 150 mm rotameters. Flow capacities shown are based on calibrations with air and water at 21°C (70°F) and 14.7 psia. For other fluids or conditions, the following equations provide rough approximations.


GAS

Qa = Qg x [G x (Ta/T0) x (P0/Pa)]1/2


where:

Qa = equivalent air flow scale reading at Standard Conditions

Qg = true flow of metered gas

G = specific gravity of metered gas

Ta = absolute temperature at flow conditions (Kelvin or Rankine)

T0 = absolute temperature at standard conditions (294 K or 530°R)

Pa = pressure at flow conditions (psia)

P0 = pressure at Standard Conditions (14.7 psia)


LIQUID

Qw = QL x [(df - dw)/(df - dL)) x (dL/dw)]1/2


where:

Qw = equivalent water flow capacity at Standard Conditions

QL = max flow of metered liquid

df = density of float selected

   Glass: 2.53 g/ml

   Sapphire: 3.98 g/ml

   Stainless Steel: 8.04 g/ml

   Carboloy: 14.98 g/ml

   Tantalum: 16.58 g/ml

dL = density of metered liquid, g/ml

dw = density of water at Standard Conditions (1.0 g/ml)

(note: does not include viscosity effect)