来宝网Logo

热门词:生物显微镜 水质分析仪 微波消解 荧光定量PCR 电化学工作站 生物安全柜

现在位置首页>自动化仪器仪表>传感器/变送器>温度传感器> M12TX-PT100-1/4-9_M12TX-PT100-1/4-10_M12TX-PT100-1

商品编号:74883
分享

M12TX-PT100-1/4-9_M12TX-PT100-1/4-10_M12TX-PT100-1

价    格:询价

产    地:美国更新时间:2021/9/10 14:45:14

品    牌:欧米茄型    号:M12TX

状    态:正常点击量:897

15221105376
联系我时,请说明是在来宝网上看到的,谢谢!

上海茂培电子科技有限公司

联 系 人:范先生

电     话:15221105376

传     真:

等     级: (第 9年)

性     质:贸易型,

联系我时请说在来宝网上看到的,谢谢!




M12TX-PT100-1/4-9_M12TX-PT100-1/4-10_M12TX-PT100-1/4-12温度传感器OMEGA

可编程发送器


成型连接器(IP67)


M12连接


4至20 mA输出=0至150°C默认设置


-50至500°C(-58至932°F)全可编程范围

Programmable Transmitter

Molded Connector (IP67)

M12 Connection

4 to 20 mA Output = 0 to 150°C

 Default Setting

-50 to 500°C (-58 to 932°F) Full Programmable Range

The M12TX 100Ω RTD sensor with built-in transmitter can be programmable by a computer. The configuration kit required for programming is M12TX-CONFIG. This unique probe is ideal for areas with space limitations where traditional head connections are too large to fit. The M12 thread design offers a secure industrial connection. The temperature Transmitter is designed with mineral insulated cable probe (MgO) which allows the probe to have a minimum bend radius of three times the diameter.



产品参数

M12TX-PT100-1/4-9_M12TX-PT100-1/4-10_M12TX-PT100-1/4-12温度传感器OMEGA

Programmable Transmitter

Molded Connector (IP67)

M12 Connection

4 to 20 mA Output = 0 to 150°C Default Setting

-50 to 500°C (-58 to 932°F) Full Programmable Range

The M12TX 100Ω RTD sensor with built-in transmitter can be programmable by a computer. The configuration kit required for programming is M12TX-CONFIG. This unique probe is ideal for areas with space limitations where traditional head connections are too large to fit. The M12 thread design offers a secure industrial connection. The temperature Transmitter is designed with mineral insulated cable probe (MgO) which allows the probe to have a minimum bend radius of three times the diameter.


Specifications

Body: Thermoplastic material

Probe Type: Compact mineral insulated cable AISI 316L SS

Probe Minimum Bending Radius: Three-times the outer diameter [except the sensing tip which length is 30 mm (1.2")]

Connection: M12x1 male in accordance to VDE0627 (output connection)

Sensor: RTD Pt100 class A up to 300°C in accordance to IEC751

Range: -50 to 500°C (-58 to 932°F)

Tmax. Electronic: 80°C (176°F)

Sensor Break Monitoring:

    Selectable: Upscale (>21.0 mA) or downscale (<3.6 mA) action

    Sensor Short-Circuit: Fixed to downscale (<3.6 mA) action

Output

Signal: 4 to 20 mA

Permissible load: 700 @ 24 Vdc [RLo= (Vsupply-8.5)/0.020]

Response Time (90%): <50 ms

Isolation In-Out: Non-isolated

Power Supply: 8.5 to 32 Vdc (polarity protected)

Environment Conditions

Temperature: -40 to 80°C (-40 to 176°F) for plastic body

Relative Humidity: 0 to 100%

EMC: In accordance to EN 61326

Degree of Protection: IP65 and IP67 in accordance to IEC60529

Accuracy

Transmitter: Maximum of ±0.2°C or ±0.2% of span

Sensor: Class A in accordance to IEC751

Temperature Influence [Deviation from 20°C (68°F)]: Maximum of ±0.3°C/25°C (77°F) or ±0.3% of span/25°C (77°F)

Supply Voltage Influence: Negligible

Range Configurations: It is possible set the input temperature range (span) by the M12TX-CONFIG configuration kit (PC with OS Windows? required)

Zero Adjustments: Any value between -50 to 50°C

Minimum Span: 50°C (122°F) [if the zero value is set between one of these values: -40°C (-40°F), -20°C (-4°F), 0°C (32°F), 20°C (68°F), 40°C (104°F), the minimum span is 20°C (68°F) rather than 50°C (122°F)]

Sensor Error Compensation: Over 2 points (maximum 1% of span)

Factory Setting: 0 to 150°C/sensor break >21 mA (upscale)

Response Time: Diameter 3 mm <3.5 second, diameter 6 mm <13 second (test in water to IEC751- time for reaching 63.2% of instantaneous temperature change) three times the diameter.



产品介绍

M12TX-PT100-1/4-9_M12TX-PT100-1/4-10_M12TX-PT100-1/4-12温度传感器OMEGA

范围配置:可以通过m12tx-config配置工具包(需要配备OS Windows?的PC)设置输入温度范围(范围)。


零位调整:在-50到50°C之间的任何值


最小跨距:50°C(122°F)【如果零值设置在以下值之一之间:-40°C(-40°F)、-20°C(-4°F)、0°C(32°F)、20°C(68°F)、40°C(104°F),最小跨距为20°C(68°F)而不是50°C(122°F)】


传感器误差补偿:超过2点(最大跨度的1%)


出厂设置:0至150°C/传感器断开>21 mA(高档)


响应时间:直径3 mm<3.5秒,直径6 mm<13秒(根据IEC751标准在水中试验-达到瞬时温度变化63.2%的时间)为直径的三倍。




猜你喜欢