英国Sherwood氯离子分析仪M926\M926S

价    格:询价

产    地:英国更新时间:2019/7/15 13:41:34

品    牌:Sherwood型    号:M926\M926S

状    态:正常点击量:1441

18253171090
联系我时,请说明是在来宝网上看到的,谢谢!

济南欧莱博科学仪器有限公司

联 系 人:李经理

电     话:18253171090

传     真:

等     级: (第 4年)

性     质:生产型,

联系我时请说在来宝网上看到的,谢谢!




使用方便, 自动化操作.
很高的灵敏度.
很高的选择性.


产品参数

               Model 926 Model 926S
样品量      500 microlitres 20 microlitres
分析范围: 10-999mg Cl/l; 2-165mg%Salt 10-299mmolCl/l
分析时间:  35 sec @ 200mgl/l 35 sec@ 100mmolCl/l
准确度:      SD <1% @ 0.1% Cl Std Solution +-2% @ NQCS 100mmol/l

线性:         1% in range 10-999mgCl/l 1% 10-299 mmolCl/l
重现性:    CV for 20 replicates
<1.5% @ 200mgCl/l <1.5% @ 100mmolCl/l


产品介绍

KEY APPLICATION AREAS
The measurement of Salt (Sodium Chloride) concentration in the Food and Dairy Industry is universal. The ability to read direct in Salt concentration makes the SHERWOOD Model 926 the instrument of choice for manufacturers and food analysts around the world. 
Salt and salinity also represents serious contamination in Industrial processes and the sensitivity of the coulometric method enables measurement in the ppm range in boiler feed water, polymer washes. bore hole slurries, as well as soils analysis and others. The 926 has proven itself to be the method of choice for many desalination studies and general quality control of sample lines. 
Clinical measurement of the chloride ion is essential for monitoring of electrolyte balance of patients as it plays, in combination with sodium and potassium, an important part in maintaining water balance between cells, plasma and interstitial fluid. The Model 926s is also widely used in the confirmation of the diagnosis of Cystic Fibrosis. The model 926S variant can accept 20 microlitre samples. 
Sherwood Scientific has method notes for these and other Applications. Pharmaceutical Production, Cement Manufacture, Photography Veterinary. 
 
PRINCIPLES OF METHOD
The main difficulties with the classic method were that silver nitrate reagent would change on storage due to a photochemical reaction and that the titration end point is visually weak. The reagent would have to be standardized requiring a highly skilled analytical technician 
The 926 method is based on a coulometric titration and is an absolute method where the reagent, the silver ions, is precisely and quantitatively generated at the time of the analysis by passing a constant current between donor electrodes. The end point is detected by the use of sensing electrodes which measure the change in solution conductivity which occurs when excess silver ions are present in solution. 