英国HEL快速筛选量热仪TSu

价    格:询价

产    地:英国更新时间:2019/7/15 13:41:34

品    牌:HEL型    号:TSu

状    态:正常点击量:2210

18253171090
联系我时,请说明是在来宝网上看到的,谢谢!

济南堃康医疗器械有限公司

联 系 人:李经理

电     话:18253171090

传     真:

等     级: (第 4年)

性     质:生产型,

联系我时请说在来宝网上看到的,谢谢!




The Thermal Screening Unit (TSu) has been developed for the first step in reaction hazards assessment, a more comprehensive solution than traditional screening by DSC/DTA .
For effective reaction hazards screening, two critical pieces of data need to be determined: the 'onset' temperature for the reaction and the pressure generated by the runaway.
The TSu offers you the simplicity of operation associated with DSC and DTA, whilst providing pressure data. In addition, by utilising sample cells of up to 8ml, representative samples can be tested.

主要特点
Alternative to DSC/DTA as a primary thermal hazard screening method
Wide temperature range
Simultaneous pressure and temperature data
Rapid sample turn around
Representative sample
Low running costs


产品参数

Operating Modes:
The standard procedure involves heating of the sample at a user defined rate (typically 0.5 to 5°C/minute) while both the temperature and pressure are recorded. Deviation from linearity (e.g. a sharp rise) indicates the 'onset' of an exotherm. The subsequent rise shows the severity of hazard directly without the need for any calculation.
Additional operating modes include isothermal, dual scan and soak-and-scan are available.

Operating Range:
Temperature from ambient to 400°C, resolution of 0.01°C. Pressure transducers available in several ranges, typically up to 200 bar.

Test Cells:
Spherical test cells of approximately 8ml in stainless steel, hastelloy, and other alloys are available as well as in glass. This large volume allows for truly representative samples, and the study of liquids, solids and reaction mixtures.

Running Experiments:
winISO has been developed in house to provide a logical and simple interface for development of multiple step processes. It runs under Windows software. During operation all conditions are displayed numerically and on a trend graph. All steps can be modified during operation. Modifications are fully documented in the output file.

Applications:
Thermal stability of feeds, intermediates and products.
Test liquids and solids.
Evaluate long term exposure at elevated temperatures (isothermal or "soak" test).
Evaluate safe operating and storage temperatures, as well as the consequences of a runaway reaction.

Key Data Produced:
Onset temperature of exotherm.
Rate of temperature rise.
Rate of pressure rise.
Maximum temperature and pressure.
Time from exotherm initiation to maximum rate.

Multiple Units:
Several TSus can be used in parallel, running from the same computer. This is a cost effective method of screening large numbers of samples. The compact design of the instrument makes it easy to fit several units into a standard fume cupboard.
 